Important: Follow instructions that came with inhaler (MDI).
![]() |
(1) Remove cap(s).
![]() |
(2) Look for foreign objects.
![]() |
(3) Insert inhaler (MDI) into back.
![]() |
(4) Apply mask to face or put mouthpiece into mouth.
![]() |
(5) Depress inhaler (MDI) at beginning of slow inhalation. If using mask, maintain good seal for 2-3 breaths after depressing inhaler.
![]() |
(6) Slow down inhalation if you hear the FLOWSIGnal® Whistle sound.
![]() |
(7) Wait 30 seconds before repeating steps 4-6 as prescribed.
![]() |
![]() |
Clean AeorChamber Plus™ VHC only as per instructions before first use, then weekly.
![]() |
Remove back piece only. (Do not remove mask)
![]() |
Soak both parts for 15 minutes in lukewarm water with liquid detergent. Agitate gently.
![]() |
Rinse in clean water.
![]() |
Shake out excess water. Do not rub dry.
![]() |
Let air dry in vertical position.
![]() |
Replace back piece when unit is completely dry and ready for use.
Notes
— VHC--Valved Holding Chamber
— Do not leave AeroChamber Plus™ VHC unattended with children. This is a medical device, not a toy.
— Do not disassemble the product at any time (even for cleaning, except for flexible back piece).
— This product contains no latex.
Cautions
— Product should be replaced if valves are damaged or missing.
— If unsure on how to use this product, talk to your physician or pharmacist.
— Administer one puff at a time.
— Rx Only
Indications for Use: The AeroChamber Plus™ Valved Holding Chamber (VHC) is intended to be used by patients who are under the care or treatment of a licensed health care professional or physician. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, prescribed by a physician or health care professional. The intended environments for use include the home, hospitals and clinics.
Distributed by:
Forest Pharmaceuticals, Inc.
Subsidiary of Forest Laboratories, Inc. St. Louis, MO 63045
Monaghan Medical Corporation, P.O. Box 2805, Plattsburgh, NY 12901
© 1999, 2001 Monaghan Medical Corporation
Patents 5,042,467; 5,645,049; 5,848,588; 5,988,160 and Patents Pending
Printed in USA RMC 5263 Revision: 02/01